To output harmonic, Overview, To output harmonic -30 – RIGOL DG1000Z Series User Manual

Page 80: Overview -30, 2 sin( ) 2 sin( ) 2 sin( )

Advertising
background image

RIGOL

Chapter 2 Front Panel Operations

2-30

DG1000Z User’s Guide

To Output Harmonic

DG1000Z can be used as a harmonic generator to output harmonic with specified

order, amplitude and phase. It is usually used in the test of harmonic detector device

or harmonic filter device. This section introduces how to configure the generator to

output harmonic.

Overview

According to Fourier transform, time domain waveform is

the superposition of a

series of sine waveforms as shown in the equation below:

......

)

2

sin(

)

2

sin(

)

2

sin(

)

(

3

3

3

2

2

2

1

1

1

+

+

+

+

+

+

=

ϕ

π

ϕ

π

ϕ

π

t

f

A

t

f

A

t

f

A

t

f

Generally, component with

1

f

frequency is called fundamental waveform,

1

f

is

fundamental waveform frequency,

1

A

is fundamental waveform amplitude,

1

ϕ

is

fundamental waveform phase. The frequencies of the other components (called

harmonics) are all integral multiples of the fundamental waveform frequency.

Components whose frequencies are odd multiples of the fundamental waveform

frequency are called odd harmonics and components whose frequencies are even

multiples of the fundamental waveform frequency are called even harmonics.


DG1000Z can output up to 8th harmonic. After selecting CH1 or CH2, press Sine

Harm to select “On” and press Harmonic Para to enter the harmonic setting menu.

You can set the type of harmonic, specify the highest order of harmonic and set the

amplitude and phase of each order of harmonic. If you want to set various

parameters of fundamental waveform, please set the parameters of Sine.


After finishing harmonic parameters setting, press Output1 and the backlight turns

on, the instrument outputs the specified harmonic from the corresponding output

terminal.