To set leading/trailing edge time (pulse), To set leading/trailing edge time (pulse) -13 – RIGOL DG1000Z Series User Manual

Page 63

Advertising
background image

Chapter 2 Front Panel Operations

RIGOL

DG1000Z User’s Guide

2-13

Users can also use the direction keys and knob to set the parameter value: use the

direction keys to move the cursor to select the digit to be edited and then rotate the

knob to change the number.

To Set Leading/Trailing Edge Time (Pulse)

Leading (rising) edge time is defined as the duration of the pulse amplitude rising

from 10% to 90% threshold, while Trailing (falling) edge time is defined as the

duration of the pulse amplitude moving down from 90% to 10% threshold (as shown

in the figure below).

The setting range of leading/trailing edge time is limited by the currently specified

pulse width limit (as shown in the formula below). DG1000Z will automatically adjust

the edge time to match the specified pulse width if the value currently set exceeds

the limit value.

Leading/Trailing Edge Time ≤ 0.625 × Pulse Width

Press Leading (Trailing) to highlight “Leading” (“Trailing”). Use the numeric

keyboard to input the desired value and then select the desired unit from the pop-up

menu.
 The Leading/Trailing Edge Time units available are sec, msec, μsec and nsec.
 Leading edge time and trailing edge time are independent from each other and

users can set them separately.

Users can also use the direction keys and knob to set the parameter value: use the

direction keys to move the cursor to select the digit to be edited and then rotate the

knob to change the number.

10%

90%

50%

t

Fall

t

Rise

Pulse Width

Pulse Period

Advertising