To Set Fundamental Waveform Parameters

DG1000Z allows users to set various fundamental waveform parameters such as

frequency, period, amplitude, DC offset voltage, high level, low level and start phase.

It also supports align phase operation. Please refer to introductions in To Output

Basic Waveform to set above fundamental waveform parameters.

To Set Harmonic Order

The highest order of harmonic output from DG1000Z can not be greater than this

setting value.

After entering the harmonic setting menu, press Order (at this point, “Order” on the

screen is highlighted) and use the numeric keyboard or knob to input the desired

harmonic order.
 The range is limited by the maximum output frequency of the instrument and

current fundamental waveform frequency.

 Range: 2 to maximum output frequency of the instrument÷current

fundamental waveform frequency (must be integer).

 The maximum is 8.

To Select Harmonic Type

DG1000Z can output even harmonic, odd harmonic, all orders of harmonic and

user-defined orders of harmonic. After entering the harmonic setting menu, press

Type to select the desired harmonic type.

1. Even

Press this key and the instrument will output fundamental waveform and even

harmonics.

2. Odd

Press this key and the instrument would output fundamental waveform and odd

harmonics.

3. All

Press this key and the instrument would output fundamental waveform and all

the harmonics in order.

4. User

Press this key and the instrument would output the user-defined orders of

harmonics. The highest order is 8.