Replacing Electronics Module or PROM,

Continued

Procedure, continued

Table 28

Replacing Electronics Module or PROM, Continued

Step

Action

3

Release retaining clip and unplug flex -tape and power connectors
from Main PWA underneath module. Unplug temperature input
connector from RTD measurement (Daughter) PWA underneath
module. Loosen two captive mounting screws on top of module, and
then carefully pull module from housing.

22365

Connectors

End-cap lock

Electronics

Module

Flex-Tape
and Power

Temperature Input
Connector

4

Remove screw holding molding/retaining clip to Main PWA and
remove molding/retaining clip from Main PWA.

Continued on next page