Operating Instructions

Model 6200E Instruction Manual

114

M6200E Rev: A1


Each analyzer in the multidrop chain must have:

• One Teledyne Analytical Instruments option 62 installed.

• One 6’ straight-through, DB9 male Æ DB9 Female cable (Teledyne Analytical Instruments

P/N WR0000101) is required for each analyzer.

To set up the network, for each analyzer:

1. Turn the analyzer on and change its ID code (see Section 6.10.1) to a unique 4-digit number.

2. Remove the top cover (see Section 3.1) of the analyzer and locate JP2 on the multidrop PCA

(see Figure 6-12)

3. Make sure that the jumpers are in place connection pins 9

↔

10 and 11

↔

12.

4. If the analyzer is to be the last instrument on the chain, make sure a jumper is in place

connecting pins 21

↔

22.

5. If you are adding an analyzer to the end of an already existing chain, don’t forget to remove

JP2, pins 21

↔

22 on the multidrop PCA on the analyzer that was previous the last instrument

in the chain.

6. Close the instrument.

7. Using straight-through, DB9 male Æ DB9 Female cables, interconnect the host and the

analyzers as shown in Figure 6-13.

NOTE:

Teledyne Analytical Instruments recommends setting up the first link, between the Host

and the first analyzer and testing it before setting up the rest of the chain.