Operating Instructions
Model 6200E Instruction Manual
136
M6200E Rev: A1
6.11.2.11. HOLDOFF Feature
The iDAS HOLDOFF feature allows to prevent data collection during calibrations and during the
DAS_HOLDOFF period enabled and specified in the VARS (Section 6.8). To enable or disable the
HOLDOFF, follow the instruction shown in section 6.11.2.2 then press
:
Edit Data Channel Menu
From the DATA ACQUISITION menu
SETUP X.X
0) CONC: ATIMER, 2, 4032, R
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
EXIT
SETUP X.X
CAL HOLD OFF:ON
SET>
EDIT
EXIT
ENTR accepts the new
setting and returns to the
previous menu.
EXIT ignores the new setting
and returns to the previous
menu.
SETUP X.X
CAL HOLD OFF:ON
ON
ENTR
EXIT
Toggle key to turn
HOLDOFF ON or OFF
Press SET> key until…