Span gas, Value – Teledyne 6200E - Sulfides Analyzer User Manual
Page 165

Model 6200E Instruction Manual
Calibration Procedures
M6200E Rev: A1
165
Step Two: Set the expected H
2
S
span gas
value:
This sequence causes the
analyzer to prompt for the
expected H
2
S span
concentration.
M-P CAL
H2S SPAN CONC: 450.0 Conc
0 0 0
4 5 .0
ENTR
EXIT
M-P CAL RANGE = 500.000 PPB
H2S
=XXX.X
< TST TST > ZERO
CONC
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST >
CAL
SETUP
The H
2
S span concentration
values automatically default
to 450.0 Conc.
To change this value to the
actual concentration of the
span gas, enter the number
by pressing the key under
each digit until the expected
value appears.
The span gas concentration
should always be 90% of the
selected reporting range
EXAMPLE
Reporting range = 800 ppb
Span gas conc.= 720 ppb
EXIT ignores the new setting
and returns to the previous
display.
ENTR accepts the new setting
and returns to the
previous display..