Calibration Procedures

Model 6200E Instruction Manual

176

M6200E Rev: A1

7.9.1. Autocal of instruments in INDEPENDENT or AUTO

Reporting Range Modes

If the analyzer is being operated in IND or AUTO Range mode, then the High and Low ranges
must be specified as part of the Auto Cal set up. This parameter appears at the end of the

programming sequences after the CALIBRATE: ON/OFF parameter is set. For example:

SETUP X.X

RANGE TO CAL: LOW


<SET

EDIT

EXIT

EXIT returns

to the SETUP

Menu

SETUP X.X

CALIBRATE: ON


<SET

SET>

EDIT

EXIT

SETUP X.X

SEQ 2) ZERO–SPAN, 2:00:30


PREV NEXT

MODE

SET

EXIT

Sequence

Delta Time

MODE Delta Days

Follow standard AutoCal programming process to this

point, then …

SETUP X.X

RANGE TO CAL: LOW

LOW

HIGH

ENT

R EXIT

SETUP X.X

RANGE TO CAL: HIGH

LOW HIGH

ENTR

EXIT

SETUP X.X

RANGE TO CAL: HIGH


<SET

EDIT

EXIT

7.9.2. Autocal of instruments in Multigas Measurement Mode

If the analyzer is being operated in multigas measurement mode, autocal can only be

programmed to calibrate the instrument with one of the available gases.

NOTE

Make sure that the gas for which the AutoCal sequence has been programmed is being

supplied to the analyzer at the time the autocal program initiates.

If the wrong gas is being supplied to the instrument, the analyzer will be mis-calibrated.