Teledyne 6200E - Sulfides Analyzer User Manual
Page 176

Calibration Procedures
Model 6200E Instruction Manual
176
M6200E Rev: A1
7.9.1. Autocal of instruments in INDEPENDENT or AUTO
Reporting Range Modes
If the analyzer is being operated in IND or AUTO Range mode, then the High and Low ranges
must be specified as part of the Auto Cal set up. This parameter appears at the end of the
programming sequences after the CALIBRATE: ON/OFF parameter is set. For example:
SETUP X.X
RANGE TO CAL: LOW
<SET
EDIT
EXIT
EXIT returns
to the SETUP
Menu
SETUP X.X
CALIBRATE: ON
<SET
SET>
EDIT
EXIT
SETUP X.X
SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT
MODE
SET
EXIT
Sequence
Delta Time
MODE Delta Days
Follow standard AutoCal programming process to this
point, then …
SETUP X.X
RANGE TO CAL: LOW
LOW
HIGH
ENT
R EXIT
SETUP X.X
RANGE TO CAL: HIGH
LOW HIGH
ENTR
EXIT
SETUP X.X
RANGE TO CAL: HIGH
<SET
EDIT
EXIT
7.9.2. Autocal of instruments in Multigas Measurement Mode
If the analyzer is being operated in multigas measurement mode, autocal can only be
programmed to calibrate the instrument with one of the available gases.
NOTE
Make sure that the gas for which the AutoCal sequence has been programmed is being
supplied to the analyzer at the time the autocal program initiates.
If the wrong gas is being supplied to the instrument, the analyzer will be mis-calibrated.