Model 6200E Instruction Manual

Theory Of Operation

M6200E Rev: A1

227

10.4.9. Status LEDs & Watch Dog Circuitry

Z ero/S pan and IZ S O ptions
S am ple/C al V alve

IZ S O ption
P erm eation T ub e H eater

D ark S hutter

S a m ple C ham b er

H eater

I

2

C

W atchdog LE D

Zero/S pan and IZS O ptions
Ze ro/S pa n V alve

H

2

S

Æ SO

2

co nverter hea te r

S O

2

/H

2

S valve

Figure 10-17: Relay Board Status LED Locations

Thirteen LEDs are located on the analyzer’s relay board to indicate the status of the analyzer’s
heating zones and valves as well as a general operating watchdog indicator. Table 10-2 shows the
states of these LEDs and their respective functionality.

Table 10-2

:

Relay Board Status LED’s

LED

COLOR

FUNCTION

STATUS WHEN LIT

STATUS WHEN UNLIT

D1

RED

Watchdog circuit

Cycles On/Off every 3 seconds under control of the CPU.

D2 YELLOW

Sample chamber

heater

HEATING NOT

HEATING

D3 YELLOW

H

2

S Æ SO

2

converter

heater

HEATING NOT

HEATING

D4

YELLOW

Unused

N/A

N/A

D5 YELLOW

IZS heater Perm.

Tube (option)

HEATING NOT

HEATING

D6

YELLOW

Unused

N/A

N/A

D7 GREEN

Sample/Cal Valve

(option)

Valve open to zero/span

valve.

Valve open to sample inlet

D8 GREEN

Zero/Span Valve
(option)

Valve open to zero gas inlet

Valve open to span gas inlet

D9 GREEN

SO

2

/H

2

S valve

Gas stream bypasses H

2

S Æ

SO

2

converter. Analyzer

measuring SO

2

Valve open to H

2

S Æ SO

2

converter. Analyzer
measuring H

2

S.

D10

GREEN

Unused

N/A

N/A

D11

GREEN

UV Lamp Shutter

Shutter open

Shutter closed

D12-14

GREEN

Unused

N/A

N/A