Model 6200E Instruction Manual
Operating Instructions
M6200E Rev: A1
65
The default status of the calibration password is OFF. To enable the calibration password press:
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SAMPLE RANGE = 500.0 PPB H2S =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS
CLK MORE
EXIT
SETUP X.X
CAL. PASSWORD ENABLE: OFF
OFF
ENTR
EXIT
SETUP X.X
PASSWORD ENABLE:
ON
ON
ENTR
EXIT
SETUP X.X
PASSWORD ENABLE: ON
ON
ENTR
EXIT
ENTR accepts
the change
EXIT ignores
the change
Toggles
password
status On/Off
ENTR accepts
displayed
password value
EXIT returns to
SAMPLE display
CAL. PASSWORD
default state is
OFF
If the calibration password (101) is enabled, the following keypad sequence will be required to
enter one of the calibration modes:
SAMPLE
ENTER SETUP PASS : 0
0 0 0
ENTR
EXIT
SAMPLE RANGE = 500.0 PPB H2S =X.XXX
< TST TST >
CAL CALZ CALS
SETUP
M-P CAL RANGE = 500.0 PPB H2S =X.XXX
< TST TST > ZERO
CONC
EXIT
Prompts
password
number
SAMPLE
ENTER SETUP PASS : 0
1 0 1
ENTR
EXIT
Press
individual
keys to set
101
Continue calibration process …