Model 6200E Instruction Manual
Getting Started
M6200E Rev: A1
39
STEP THREE: Perform the zero/span calibration procedure:
M-P CAL
H2S STB=X.XXX PPB
SO2 =X.XXX
< TST TST >
ENTR
CONC
EXIT
M-P CAL
H2S STB=X.XXX PPB
SO2 =X.XXX
< TST TST >
CAL
SETUP
M-P CAL
H2S STB=X.XXX PPB
SO2 =X.XXX
< TST TST >
ZERO
CONC
EXIT
EXIT returns to the main
SAMPLE display
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
M-P CAL
H2S STB=X.XXX PPB
H2S =X.XXX
< TST TST >
SPAN
CONC
EXIT
M-P CAL RANGE = 500.0 PPB H2S =X.XXX
< TST TST >
ENTR
SPAN CONC
EXIT
M-P CAL
RANGE = 500.0 PPB H2S =X.XXX
< TST TST > ENTR CONC
EXIT
Wait until H2S STB
falls below 0.5 ppb.
This may take several
minutes.
The SPAN key now
appears during the
transition from zero to
span.
You may see both keys.
If either the ZERO or
SPAN buttons fail to
appear see Section 11
for troubleshooting tips.
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST
TST
>
CAL
SETUP
SAMPLE
H2S STB=X.XXX PPB
H2S =X.XXX
< TST TST >
CAL
SETUP
Set the Display to show the H2S
STB test function.
This function calculates the
stability of the H
2
S
x
measurement
Press ENTR to changes the
OFFSET & SLOPE values for the
SO
2
measurements.
Press EXIT to leave the calibration
unchanged and return to the
previous menu.
ACTION:
Allow span gas to enter the sample port at the
rear of the instrument.
The value of
H2S STB may jump
significantly.
Wait until it falls back
below 0.5 ppb.
This may take several
minutes.
Press ENTR to change the offset &
slope values for the H
2
S
measurements.
Press EXIT to leave the calibration
unchanged and return to the
previous menu.
The Model 6200E analyzer is now ready for operation.