Teledyne 6200E - Sulfides Analyzer User Manual

Page 39

Advertising
background image

Model 6200E Instruction Manual

Getting Started

M6200E Rev: A1

39

STEP THREE: Perform the zero/span calibration procedure:

M-P CAL

H2S STB=X.XXX PPB

SO2 =X.XXX

< TST TST >

ENTR

CONC

EXIT

M-P CAL

H2S STB=X.XXX PPB

SO2 =X.XXX

< TST TST >

CAL

SETUP

M-P CAL

H2S STB=X.XXX PPB

SO2 =X.XXX

< TST TST >

ZERO

CONC

EXIT

EXIT returns to the main

SAMPLE display

ACTION:

Allow zero gas to enter the sample port at the

rear of the instrument.

M-P CAL

H2S STB=X.XXX PPB

H2S =X.XXX

< TST TST >

SPAN

CONC

EXIT

M-P CAL RANGE = 500.0 PPB H2S =X.XXX

< TST TST >

ENTR

SPAN CONC

EXIT

M-P CAL

RANGE = 500.0 PPB H2S =X.XXX

< TST TST > ENTR CONC

EXIT

Wait until H2S STB
falls below 0.5 ppb.

This may take several

minutes.

The SPAN key now

appears during the

transition from zero to

span.

You may see both keys.

If either the ZERO or

SPAN buttons fail to

appear see Section 11

for troubleshooting tips.

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X

< TST

TST

>

CAL

SETUP

SAMPLE

H2S STB=X.XXX PPB

H2S =X.XXX

< TST TST >

CAL

SETUP

Set the Display to show the H2S

STB test function.

This function calculates the

stability of the H

2

S

x

measurement

Press ENTR to changes the

OFFSET & SLOPE values for the

SO

2

measurements.

Press EXIT to leave the calibration

unchanged and return to the

previous menu.

ACTION:

Allow span gas to enter the sample port at the

rear of the instrument.

The value of

H2S STB may jump

significantly.

Wait until it falls back

below 0.5 ppb.

This may take several

minutes.

Press ENTR to change the offset &

slope values for the H

2

S

measurements.

Press EXIT to leave the calibration

unchanged and return to the

previous menu.

The Model 6200E analyzer is now ready for operation.