Level 1 calibrations versus level 2 checks – Teledyne 6200E - Sulfides Analyzer User Manual

Page 183

Advertising
background image

Model 6200E Instruction Manual

EPA Protocol Calibration

M6200E Rev: A1

183

8.2. Level 1 Calibrations versus Level 2 Checks

Essential to quality assurance are scheduled checks for verifying the operational status of the
monitoring system. The operator should visit the site at least once each week. It is recommended

Level 1 zero and span check conducted on the analyzer every two weeks. Level 2 zero and span
checks should be conducted at a frequency desired by the user. Definitions of these terms are

given in Table 8-4.

In addition, an independent precision check between 0.08 and 0.10 ppm must be carried out at

least once every two weeks. Table 8-3 summarizes the quality assurance activities for routine
operations. A discussion of each activity appears in the following sections.

To provide for documentation and accountability of activities, a checklist should be compiled and

then filled out by the field operator as each activity is completed.

Table 8-4: Definition of Level 1 and Level 2 Zero and Span Checks

(from Section 2.0.9 of Q.A. Handbook for Air Pollution Measurement Systems)

LEVEL 1 ZERO AND SPAN CALIBRATION
A Level 1 zero and span calibration is a

simplified, two-point analyzer calibration used
when analyzer linearity does not need to be
checked or verified. (Sometimes when no

adjustments are made to the analyzer, the Level
1 calibration may be called a zero/span check, in
which case it must not be confused with a Level 2
zero/span check.) Since most analyzers have a

reliably linear or near-linear output response with
concentration, they can be adequately calibrated
with only two concentration standards (two-point
concentration). Furthermore, one of the
standards may be zero concentration, which is
relatively easily obtained and need not be
certified. Hence, only one certified concentration
standard is needed for the two-point (Level 1)
zero and span calibration. Although lacking the
advantages of the multipoint calibration, the two-
point zero and span calibration--because of its
simplicity--can be (and should be) carried out
much more frequently. Also, two-point
calibrations are easily automated. Frequency
checks or updating of the calibration relationship

with a two-point zero and span calibration
improves the quality of the monitoring data by
helping to keep the calibration relationship more

closely matched to any changes (drifts) in the
analyzer response.

LEVEL 2 ZERO AND SPAN CHECK
A Level 2 zero and span check is an "unofficial" check

of an analyzer's response. It may include dynamic
checks made with uncertified test concentrations,
artificial stimulation of the analyzer's detector,

electronic or other types of checks of a portion of the
analyzer, etc.
Level 2 zero and span checks are not to be used as a
basis for analyzer zero or span adjustments,
calibration updates, or adjustment of ambient data.
They are intended as quick, convenient checks to be
used between zero and span calibrations to check for

possible analyzer malfunction or calibration drift.
Whenever a Level 2 zero or span check indicates a
possible calibration problem, a Level 1 zero and span
(or multipoint) calibration should be carried out

before any corrective action is taken.
If a Level 2 zero and span check is to be used in the
quality control program, a "reference response" for
the check should be obtained immediately following a
zero and span (or multipoint) calibration while the
analyzer's calibration is accurately known.
Subsequent Level 2 check responses should then be
compared to the most recent reference response to
determine if a change in response has occurred. For
automatic Level 2 zero and span checks, the first

scheduled check following the calibration should be
used for the reference response. It should be kept in
mind that any Level 2 check that involves only part of
the analyzer's system cannot provide information

about the portions of the system not checked and
therefore cannot be used as a verification of the
overall analyzer calibration.

Advertising