Pressure calibration – Teledyne 6200E - Sulfides Analyzer User Manual
Page 98

Operating Instructions
Model 6200E Instruction Manual
98
M6200E Rev: A1
6.9.8. Pressure Calibration
A sensor at the exit of the sample chamber continuously measures the pressure of the sample gas.
This data is used to compensate the final H
2
S concentration calculation for changes in atmospheric
pressure when the instrument’s TPC feature (see Section 10.7.3) is turned on and is stored in the
CPU’s memory as the test function PRES (also viewable via the front panel).
To cause the analyzer to measure and record a value for PRES, press.
Exit at
any time
to return
to main
the
SETUP
menu
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Repeat Pressing NEXT until . .
.
DIAG
PRESSURE CALIBRATION
PREV NEXT
ENTR
EXIT
DIAG FCAL
LAMP CAL VALUE:27.20 IN-HG-A
2 7 .2 0 ENTR EXIT
ENTR accepts the
new value
EXIT ignores the new
value
The value displayed is the
current output of the UV
source reference detector
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT