3.2. Initial Operation
CAUTION
Do not look at the UV lamp while the unit is operating.
UV light can case eye damage.
Always use safety glasses
(plastic glasses will not do).
If you are unfamiliar with the M6200E theory of operation, we recommend that you read Chapter
10 before proceeding.
For information on navigating the analyzer’s software menus, see the menu trees described in
Appendix A.1.
3.2.1. Startup
After electrical and pneumatic connections are made, turn on the instrument and pump power.
The exhaust and PMT cooler fans should start. The display should immediately display a single,
horizontal dash in the upper left corner of the display. This will last approximately 30 seconds
while the CPU loads the operating system.
Once the CPU has completed this activity it will begin loading the analyzer firmware and
configuration data. During this process, string of messages will appear on the analyzer’s front
panel display: