Teledyne 6200E - Sulfides Analyzer User Manual
Page 155

Model 6200E Instruction Manual
Operating Instructions
M6200E Rev: A1
155
To assign or reset the status flag bit assignments, press:
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG ALRM
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SETUP X.
HESSEN STATUS FLAGS
<SET
SET>
EDIT
EXIT
SETUP X.X
COMMUNICATIONS MENU
ID
HESN
COM1
COM2
EXIT
SETUP X.
PMT DET WARNING: 0002
PREV NEXT
EDIT
PRNT
EXIT
Repeat pressing SET> until …
SETUP X.
SYSTEM RESET: 0000
PREV NEXT
EDIT
PRNT
EXIT
The <CH and
CH> keys move
the [ ] cursor left
and right along
the bit string.
Repeat pressing NEXT or PREV until the desired
message flag is displayed. See Table 6-27.
For xxample …
ENTR key accepts the
new settings
EXIT key ignores the
new settings
SETUP X.
SYSTEM RESET: [0]000
<CH CH>
[0] ENTR EXIT
Press the [?] key repeatedly to cycle through the available character set: 0-9
Note: Values of A-F can also be set but are meaningless.