Teledyne 6200E - Sulfides Analyzer User Manual

Page 133

Advertising
background image

Model 6200E Instruction Manual

Operating Instructions

M6200E Rev: A1

133

reduce the maximum amount of data points somewhat. In general, however, the maximum data
capacity is divided amongst all channels (max: 20) and parameters (max: 50 per channel).

The iDAS will check the amount of available data space and prevent the user from specifying too
many records at any given point. If, for example, the iDAS memory space can accommodate 375

more data records, the ENTR key will disappear when trying to specify more than that number of
records. This check for memory space may also make an upload of an iDAS configuration with

APICOM or a Terminal program fail, if the combined number of records would be exceeded. In this
case, it is suggested to either try from the front panel what the maximum number of records can

be or use trial-and-error in designing the iDAS script or calculate the number of records using the
DAS or APICOM manuals. To set the number of records for one channel from the front panel, press
SETUP-DAS-EDIT-ENTR and the following key sequence.


Edit Data Channel Menu

SETUP X.X

0) CONC: ATIMER, 2, 900


PREV NEXT INS DEL

EDIT

PRNT

EXIT

Exits to the main

Data Acquisition

menu

SETUP X.X

NAME:CONC


<SET

SET>

EDIT

PRINT

EXIT

SETUP X.X

NUMBER OF RECORDS:000


<SET SET>

EDIT

PRINT

EXIT

SETUP X.X

REPORT PERIODD:DAYS:0

0 0 0 0 0 ENTR EXIT

ENTR accepts the new

setting and returns to the

previous menu.

EXIT ignores the new setting

and returns to the previous

menu.

Toggle keys to set
number of records

(1-99999)

SETUP X.X

EDIT RECOPRDS (DELET DATA)

YES NO

NO returns to the

previous menu.

YES will delete all data

in this channel.

Press SET> key until…

From the DATA ACQUISITION menu

(see Section 6.12.2.2)

Advertising