Model 6200E Instruction Manual
Operating Instructions
M6200E Rev: A1
125
NOTE
Sending an iDAS configuration to the analyzer through its COM ports will replace the
existing configuration and will delete all stored data. Back up any existing data and the
iDAS configuration before uploading new settings.
6.11.2.1. Viewing iDAS Data and Settings
iDAS data and settings can be viewed on the front panel through the following keystroke sequence.
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
CONC
: DATA AVAILABLE
NEXT
VIEW
EXIT
SETUP X.X
PNUMTC
: DATA AVAILABLE
PREV
NEXT
VIEW
EXIT
SETUP X.X
CALDAT
: DATA AVAILABLE
PREV
VIEW
EXIT
SETUP X.X
00:00:00 S2SLP1=0.000
PV10 PREV
<PRM
PRM>
EXIT
SETUP X.X
00:00:00 S2SCN1 =0.0 PPM
PV10 PREV
NEXT NX10
<PRM
PRM>
EXIT
SETUP X.X
00:00:00 SMPFLW=000.0 cc / m
<PRM
PRM>
EXIT
SETUP X.X
DATA ACQUISITION
VIEW
EDIT
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG
DAS
RNGE
PASS CLK MORE
EXIT
VIEW KEYPAD FUNCTIONS
KEY FUNCTION
<PRM
Moves to the next Parameter
PRM>
Moves to the previous
Parameter
NX10
Moves the view forward 10
data points/channels
NEXT
Moves to the next data
point/channel
PREV
Moves to the previous data
point/channel
PV10
Moves the view back 10 data
points/channels
Keys only appear as needed
EXIT will return to the
main SAMPLE Display.
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT