Model 6200E Instruction Manual

Calibration Procedures

M6200E Rev: A1

173

7.9. Automatic Calibration/Checks (AutoCal)

The AutoCal system allows unattended, periodic operation of the zero/span valve options by using
the analyzer’s internal time of day clock. AutoCal operates by executing user-defined sequences to

initiate the various calibration modes of the analyzer and to open and close valves appropriately. It
is possible to program and run up to three separate sequences (SEQ1, SEQ2 and SEQ3). Each

sequence can operate in one of three modes or be disabled.

Table 7-2: AutoCal Modes

MODE ACTION

DISABLED

Disables the sequence

ZERO

Causes the sequence to perform a zero calibration or check

ZERO-SPAN

Causes the sequence to perform a zero and span

concentration calibration or check

SPAN

Causes the sequence to perform a span concentration

calibration or check

Each mode has seven parameters that control operational details of the sequence (Table 7-3).

Table 7-3: AutoCal Attribute Setup Parameters

Attribute Name

ACTION

Timer Enabled

Turns on the Sequence timer

Starting Date

Sequence will operate on Starting Date

Starting Time

Sequence will operate at Starting Time

Delta Days

Number of days to skip between each sequence

Delta Time

Incremental delay on each Delta Day that the sequence starts.

Duration

Duration of the sequence in minutes

Calibrate

Enable to do dynamic zero/span calibration, disable to do a cal
check only. This must be set to OFF for units used in US EPA
applications and with IZS option installed.

NOTE

The programmed STARTING_TIME must be a minimum of 5 minutes later than the real

time clock (See Section 6.6 for setting real time clock).