7.9. Automatic Calibration/Checks (AutoCal)
The AutoCal system allows unattended, periodic operation of the zero/span valve options by using
the analyzer’s internal time of day clock. AutoCal operates by executing user-defined sequences to
initiate the various calibration modes of the analyzer and to open and close valves appropriately. It
is possible to program and run up to three separate sequences (SEQ1, SEQ2 and SEQ3). Each
sequence can operate in one of three modes or be disabled.
Table 7-2: AutoCal Modes
MODE ACTION
DISABLED
Disables the sequence
ZERO
Causes the sequence to perform a zero calibration or check
ZERO-SPAN
Causes the sequence to perform a zero and span
concentration calibration or check
SPAN
Causes the sequence to perform a span concentration
calibration or check
Each mode has seven parameters that control operational details of the sequence (Table 7-3).
Table 7-3: AutoCal Attribute Setup Parameters
Attribute Name
ACTION
Timer Enabled
Turns on the Sequence timer
Starting Date
Sequence will operate on Starting Date
Starting Time
Sequence will operate at Starting Time
Delta Days
Number of days to skip between each sequence
Delta Time
Incremental delay on each Delta Day that the sequence starts.
Duration
Duration of the sequence in minutes
Calibrate
Enable to do dynamic zero/span calibration, disable to do a cal
check only. This must be set to OFF for units used in US EPA
applications and with IZS option installed.
NOTE
The programmed STARTING_TIME must be a minimum of 5 minutes later than the real
time clock (See Section 6.6 for setting real time clock).