Teledyne 6200E - Sulfides Analyzer User Manual
Page 170

Calibration Procedures
Model 6200E Instruction Manual
170
M6200E Rev: A1
STEP TWO: Perform the zero/span check.
Wait until H2S
STB falls below
0.5 ppb. This may
take several
minutes.
The value of H2S
STB may jump
significantly. Wait
until H2S STB falls
below 0.5 ppb. This
may take several
minutes.
SPAN CAL M
H2S STB=XXX.X PPB
H2S =XXX.X
< TST TST >
ZERO
SPAN
CONC
EXIT
SAMPLE
H2S STB=XXX.X PPB
H2S =XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
ZERO CAL M
H2S STB=XXX.X PPB
H2S =XXX.X
< TST TST >
ZERO
CONC
EXIT
SAMPLE
H2S STB=XXX.X PPB
H2S =XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
EXIT returns to the main
SAMPLE display
ACTION:
Record the
H
2
S readings
presented in the
upper right corner of
the display.
ACTION:
Record the
H
2
S readings
presented in the
upper right corner of
the display.
SAMPLE RANGE = 500.0 PPB
H2S =XXX.X
< TST TST >
CAL
CALZ
CALS
SETUP
Scroll to the H2S
STB test function.