Electrical connections, Figure 3-2: rear panel layout – Teledyne 6200E - Sulfides Analyzer User Manual
Page 23

Model 6200E Instruction Manual
Getting Started
M6200E Rev: A1
23
8. VENTILATION CLEARANCE: Whether the analyzer is set up on a bench or installed into an
instrument rack, be sure to leave sufficient ventilation clearance.
AREA
MINIMUM REQUIRED CLEARANCE
Back of the instrument
10 cm / 4 inches
Sides of the instrument
2.5 cm / 1 inch
Above and below the instrument.
2.5 cm / 1 inch
• Various rack mount kits are available for this analyzer. See Chapter 5 of this manual for
more information.
3.1.1. Electrical Connections:
CAUTION
Check the voltage and frequency label on the rear panel of the
instrument (See Figure 3-2) for compatibility with the local
power before plugging the M6200E into line power.
Do not plug in the power cord if the voltage or
frequency is incorrect.
CAUTION
Power connection must have functioning ground connection.
Do not defeat the ground wire on power plug.
Turn off analyzer power before disconnecting or
connecting electrical subassemblies.
Do not operate with cover off.
Figure 3-2: Rear Panel Layout