EPA Protocol Calibration
Model 6200E Instruction Manual
186
M6200E Rev: A1
calibration relationship. A minimum of three reference points and one zero point uniformly spaced
covering 0 to 90 percent of the operating range are recommended to define this relationship.
The analyzer's recorded response is compared with the known concentration to derive the
calibration relationship.
To perform a precision check, the instrument set up, sources of zero air and sample gas should
conform to those described in Section 7.2.
Follow the procedures described in section 7.2 for calibrating the zero points.
For each mid point:
SPAN CAL M
RANGE = 500.0 PPB
SO
2
X.XXX
< TST TST >
ZERO
SPAN
CONC
EXIT
Wait until
SO2 STB falls
below 0.5 ppb.
This may take
several minutes.
SAMPLE
SO2 STB =X.XXX PPB
SO2 =X.XXX
< TST TST > CAL
CALZ
CALS
SETUP
Record the SO
2
reading as
displayed on the
instrument’s front
panel
Press EXIT to
Return to the
Main SAMPLE
Display
ACTION:
Allow Calibration Gas diluted to
proper concentration for
Midpoint N+1 to enter the sample
port
ACTION:
Allow calibration gas diluted to proper concentration for
Midpoint N to enter the sample port
SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX
< TST
TST
>
CAL
SETUP
SAMPLE
SO2 STB =X.XXX PPB
SO2 =X.XXX
< TST TST >
CAL
SETUP
Set the Display to show the
SO2 STB test function.
This function calculates the
stability of the NO/NO
x
measurement
8.6. Special Calibration Requirements for Independent
Range or Auto Range
If Independent Range or Auto Range is selected, then it should be calibrated for both Range1 and
Range2 separately.
For zero and span point calibration, follow the procedure described in Section 7.2. Repeat the
procedure for both the HIGH and LOW Ranges