Teledyne 6200E - Sulfides Analyzer User Manual

Page 186

Advertising
background image

EPA Protocol Calibration

Model 6200E Instruction Manual

186

M6200E Rev: A1

calibration relationship. A minimum of three reference points and one zero point uniformly spaced
covering 0 to 90 percent of the operating range are recommended to define this relationship.

The analyzer's recorded response is compared with the known concentration to derive the
calibration relationship.

To perform a precision check, the instrument set up, sources of zero air and sample gas should
conform to those described in Section 7.2.

Follow the procedures described in section 7.2 for calibrating the zero points.

For each mid point:


SPAN CAL M

RANGE = 500.0 PPB

SO

2

X.XXX

< TST TST >

ZERO

SPAN

CONC

EXIT

Wait until

SO2 STB falls

below 0.5 ppb.

This may take
several minutes.

SAMPLE

SO2 STB =X.XXX PPB

SO2 =X.XXX

< TST TST > CAL

CALZ

CALS

SETUP

Record the SO

2

reading as

displayed on the

instrument’s front

panel

Press EXIT to

Return to the

Main SAMPLE

Display

ACTION:

Allow Calibration Gas diluted to

proper concentration for

Midpoint N+1 to enter the sample

port

ACTION:

Allow calibration gas diluted to proper concentration for

Midpoint N to enter the sample port

SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX

< TST

TST

>

CAL

SETUP

SAMPLE

SO2 STB =X.XXX PPB

SO2 =X.XXX

< TST TST >

CAL

SETUP

Set the Display to show the

SO2 STB test function.

This function calculates the

stability of the NO/NO

x

measurement

8.6. Special Calibration Requirements for Independent

Range or Auto Range

If Independent Range or Auto Range is selected, then it should be calibrated for both Range1 and

Range2 separately.

For zero and span point calibration, follow the procedure described in Section 7.2. Repeat the

procedure for both the HIGH and LOW Ranges