Teledyne 6200E - Sulfides Analyzer User Manual
Page 87

Model 6200E Instruction Manual
Operating Instructions
M6200E Rev: A1
87
To calibrate the outputs as a group, activate the ANALOG I/O CONFIGURATION MENU (see
Section 6.9.1), then press:
DIAG AIO
AUTO
CALIBRATING CONC_OUT_1
AUTO
CALIBRATING CONC_OUT_2
AUTO CALIBRATING TEST_OUTPUT
Exit at any time
to return to the
main DIAG
menu
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED:
NO
< SET
SET>
CAL
EXIT
If AutoCal has been
turned off for any
channel, the message
for that channel will be
similar to:
NOT AUTO CAL
CONC_OUT_1
Exit to return to
the I/O
configuration
menu
STARTING FROM DIAGNOSTIC MENU
(see Section 6.9.1)
If any of the channels have
not been calibrated this
message will read NO.
DIAG AIO AOUTS CALIBRATED:
YES
< SET
SET> CAL
EXIT
To automatically calibrate an single analog channel, activate the ANALOG I/O
CONFIGURATION MENU (see Section 6.9.1), then press: