TROUBLESHOOTING & REPAIR

Model 6200E Instruction Manual

264

M6200E Rev: A1

NOTE

Do not overload the PMT by accidentally setting both adjustment switches to their

maximum setting. This can cause permanent damage to the PMT.

11. Adjust the NORM PMT value with the gain potentiometer down to 800±10 mV. This is the final

very-fine adjustment.

12. Perform software span and zero calibrations (, 7.5Section 7.2, 7.4 or 7.8) to normalize the

sensor response to its new PMT sensitivity.

13. Review the slope and offset values, the slopes should be 1.000±0.300 and the offset values

should be 0±20 mV (-20 to +150 mV is allowed

11.7.

Technical Assistance

If this manual and its trouble-shooting / repair sections do not solve your problems, technical
assistance may be obtained from Teledyne Analytical Instruments, Customer Service, 16830
Chestnut St., City of Industry Ca. 91748, or Phone: 626-951-9221. Before you contact customer
service, fill out the problem report form in Appendix C, which is also available online for electronic
submission a

t http://www.teledyne-ai.com.