6.11.2.3. Trigger Events
To edit the list of data parameters associated with a specific data channel, press:
SETUP X.X
EVENT:ATIMER
<PREV NEXT>
ENTR
EXIT
Edit Data Channel Menu
SETUP X.X
0) CONC: ATIMER, 2, 4032,R
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the Main
Data Acquisition
menu
Press each key repeatedly to cycle through the
list of available trigger events.
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
EXIT
SETUP X.X
EVENT:ATIMER
<SET SET>
EDIT
EXIT
ENTR accepts the new string
and returns to the previous
menu.
EXIT ignores the new string
and returns to the previous
menu.
From the DATA ACQUISITION menu
(see Section 6.11.2.2)
6.11.2.4. Editing iDAS Parameters
Data channels can be edited individually from the front panel without affecting other data channels.
However, when editing a data channel, such as during adding, deleting or editing parameters, all
data for that particular channel will be lost, because the iDAS can store only data of one format
(number of parameter columns etc.) for any given channel. In addition, an iDAS configuration can
only be uploaded remotely as an entire set of channels. Hence, remote update of the iDAS will
always delete all current channels and stored data.
To modify, add or delete a parameter, follow the instruction shown in section 6.11.2.2 then press: