Figure 3-4: status output connector – Teledyne 6200E - Sulfides Analyzer User Manual

Page 25

Advertising
background image

Model 6200E Instruction Manual

Getting Started

M6200E Rev: A1

25

C

o

nn

ect

t

o In

ternal

G

rou

nd of

M

o

n

itor

in

g

STATUS

1 2 3 4 5 6 7 8 D

+

SY

STE

M

O

K

H

IG

H RA

NG

E

CONC

V

AL

ID

ZE

RO C

A

L

SPA

N CA

L

DI

AGNOS

T

IC

MODE

Figure 3-4: Status Output Connector

NOTE

Most PLC’s have internal provisions for limiting the current the input will draw. When

connecting to a unit that does not have this feature, external resistors must be used to

limit the current through the individual transistor outputs to ≤50mA (120 Ω for 5V

supply).

Table 3-2: Status Output Signals

REAR PANEL

LABEL

STATUS

DEFINITION

CONDITION

1

SYSTEM OK

ON if no faults are present.

2

CONC VALID

OFF any time the HOLD OFF feature is active, such as during calibration
or when other faults exist possibly invalidating the current concentration

measurement (example: sample flow rate is outside of acceptable limits).
ON if concentration measurement is valid.

3

HIGH RANGE ON if unit is in high range of the AUTO Range Mode

4

ZERO CAL

ON whenever the instrument’s ZERO point is being calibrated.

5

SPAN CAL

ON whenever the instrument’s SPAN point is being calibrated.

6

DIAG MODE

ON whenever the instrument is in DIAGNOSTIC mode

7 - 8

SPARE

D

EMITTER BUS The emitters of the transistors on pins 1-8 are bussed together.

SPARE

+

DC POWER

+ 5 VDC, 300 mA source (combined rating with Control Output, if used).

Digital

Ground

The ground level from the analyzer’s internal DC power supplies