Electrical test – Teledyne 6200E - Sulfides Analyzer User Manual
Page 96
Operating Instructions
Model 6200E Instruction Manual
96
M6200E Rev: A1
6.9.6. Electrical Test
The electrical test function creates a current, which substitutes the PMT signal, and feeds it into the
preamplifier board. This signal is generated by circuitry on the pre-amplifier board itself and tests
the filtering and amplification functions of that assembly along with the A/D converter on the
motherboard. It does not test the PMT itself. The electrical test should produce a PMT signal of
about 2000 ±1000 mV.
To activate the electrical test press the following keys.
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
DIAG
SIGNAL I / O
PREV
NEXT
JUMP
ENTR
EXIT
DIAG ELEC
RANGE = 500.000 PPB O2=X.XXX
<TST TST>
EXIT
Press NEXT until…
DIAG
ELECTRICAL TEST
PREV NEXT
ENTR
EXIT
Press TST until…
DIAG ELEC
PMT = 1732 MV
SO2=X.XXX
<TST TST>
EXIT
While the electrical test is
activated, PMT should equal:
2000 mV ± 1000 mV