Model 6200E Instruction Manual
Operating Instructions
M6200E Rev: A1
153
6.12.4.6. Hessen Protocol Gas ID
Since the M6200E can be, when the proper optional equipment is installed and operating, a
multigas instrument that measures both H
2
S and SO
2
, both of these gases are listed in the Hessen
protocol gas list. In its default state the Hessen protocol firmware assigns both gases a Hessen ID
number and actively reports both even if the instrument is only measuring one.
To change or edit these settings press:
KEY FUNCTION
<PREV
Moves to next gas entry in list
NEXT>
Moves the cursor previous gas entry in list
INS
Inserts a new gas entry into the list.
DEL
Deletes the >>>>>.
ENTR
Accepts the new setting and returns to the previous menu.
EXIT
Ignores the new setting and returns to the previous menu.
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG ALRM
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
HESSEN RESPONSE MODE :CMD
<SET
SET>
EDIT
EXIT
SETUP X.X
COMMUNICATIONS MENU
ID
HESN
COM1
COM2
EXIT
SETUP X.X
HESSEN VARIATION: TYPE 1
SET>
EDIT
EXIT
SETUP X.X
HESSEN GAS LIST
<SET
SET>
EDIT
EXIT
ENTR key
accepts the new
settings
EXIT key ignores
the new settings
SETUP X.X
SO2, 111, REPORTED
<PREV
NEXT>
INS
DEL
EDIT
PRNT EXIT
SETUP X.X
GAS TYPE SO2
<PREV
NEXT>
ENTR
EXIT
Use the PREV & NEXT keys to cycle
between the SO2 & H2S
SETUP X.X
GAS ID: 111
0 0 0
ENTR
EXIT
Use the PREV & NEXT keys to cycle
between the SO2 & H2S
SETUP X.X
REPORTED : ON
ON
ENTR
EXIT
Toggle this key to switch reporting
Between ON and OFF
Use the PREV & NEXT keys to cycle
existing entries in Hessen gas list