Manual calibration with zero/span valves, Model 6200e – Teledyne 6200E - Sulfides Analyzer User Manual
Page 164

Calibration Procedures
Model 6200E Instruction Manual
164
M6200E Rev: A1
7.4. Manual Calibration with Zero/Span Valves
Zero and Span calibrations using the Zero/Span Valve option are similar to that described in
Section 7.2, except that:
Zero air and span gas is supplied to the analyzer through the zero gas and span gas inlets rather
than through the sample inlet.
The zero and cal operations are initiated directly and independently with dedicated keys (CALZ &
CALS)
STEP ONE: Connect the sources of zero air and span gas to the respective ports on the rear panel
(Figure 3-2) as shown below.
External Zero
Air Scrubber
Filter
MODEL 701
Zero Air
Generator
Source of
SAMPLE Gas
VENT if input is pressurized
MODEL 700
Gas Dilution Calibrator
(with O
3
generator option)
VENT
MODEL
6200E
Sample
Zero Air
Exhaust
Span
Calibrated H
2
S gas
(At high concentration)
Figure 7-2: Setup for Manual Calibration with Z/S Valve Option Installed