IM 01C25R51-01E

7.1.5 Fluid Operating Range Setup

(1) Gas or Steam (excluding Category: Saturated steam),

Figure 7.9

Fluid Operating Range Setup Screen (1)

Input operating ranges and select units
Pressure and Temperature units are selectable from following.

Pressure unit: kPa abs, bar abs, psi abs, kPa gauge, bar gauge, psi gauge
Atmospheric Pressure unit: psi abs, bar abs, kPa abs
Temperature unit: degC, degF, Kelvin

Input reference condition and select units
Pressure and Temperature units are selectable from following.

Pressure unit: kPa abs, bar abs, psi abs
Temperature unit: degC, degF, Kelvin

NOTE

• After clicking the [Next] button, if a “The specified operating condition includes both gas

and liquid phases.” information message is displayed, the operating condition is near the

acceptable limits for liquid and gas.

Figure 7.10 Warning message

• Please confirm the density value on the following page (fluid physical property setup page)

and continue configuration.

• If the temperature and pressure range input on the Fluid Operating Range Setup is too wide,

there is a probability of the calculation error of density is big.

• These ranges are used only for the density and viscosity calculation.

• Ranges of Transmitter are not changed.