<7. Operation of EJXMVTool>

7-3

IM 01C25R51-01E

(3) Discharge Coefficient of the Primary Device

This parameter is active only when ‘Fix’ selected for [Type].

Input the Discharge Coefficient manually when ‘Fix’ is selected.

(4) Primary Device Gas Expansion Factor

This parameter is active only when ‘Fix’ selected for [Type].

Input the Gas Expansion Factory manually when ‘Fix’ is selected.

(5) Primary Device Sizing

Set the following values and select a unit.

Diameter of Differential Producer

unit: mm (default)/in

measuring temperature: 20 (default)

unit: degC (default), degF, Kelvin

(6) Primary Device Material

The primary device material selectable from followings.

Thermal expansion of the primary device is automatically calculated according to the

material.

Table 7.2

Primary device material

Material

Remarks

304 Stainless Steel

Default

316 Stainless Steel
Carbon Steel
Hastelloy C
Monel

(7) Pipe Sizing

Set the following values and select a unit.

Pipe inside Diameter

unit: mm (default) / in

Measuring temperature: 20 (default)

unit: degC (default), degF, Kelvin

(8) Pipe material

The pipe material selected from followings.

Thermal expansion of pipe is automatically calculated according to the material.

Table 7.3

Pipe material

Pipe Material

Remarks

304 Stainless Steel
316 Stainless Steel
Carbon Steel

Default

Hastelloy C
Monel