3 apply flow configuration, Apply flow configuration -18 – Yokogawa EJX930A User Manual

Page 134

Advertising
background image

<8. Operation of DYFMVTool>

8-18

IM 01C25R51-01E

8.2.3 Apply Flow Configuration

Figure 8.18 Apply Flow Configuration Screen

In order to download the configuration to the device, click “Apply”.
Then perform “Download to Device” from Device DTM. (Refer to section 6.2.3)
Refer to Chapter 5 and Figure 6.3.

NOTE

Following shows “Device General Parameters” which is applied to “Offline Database”.

Table 8.9 “Device General Parameters” applied to “Offline Database”

Simple Compensation Mode

Saturated steam

(Temperature)

Saturated steam

(Pressure)

Superheat steam

Gas temperature

pressure

compensation

Liquid temperature

compensation

Arithmetic Type

Applied

Applied

Applied

Applied

Applied

Bias

Applied (Always "0") Applied (Always "0") Applied (Always "0") Applied (Always "0") Applied (Always "0")

Gain

Applied (Always "1") Applied (Always "1") Applied (Always "1") Applied (Always "1") Applied (Always "1")

Bias Input 1

Applied (Always "0") Applied (Always "0") Applied (Always "0") Applied (Always "0") Applied (Always "0")

Gain Input 1

Applied (Always "1") Applied (Always "1") Applied (Always "1") Applied (Always "1") Applied (Always "1")

Bias Input 2

Applied (Always "0") Applied (Always "0") Applied (Always "0") Applied (Always "0") Applied (Always "0")

Gain Input 2

Applied (Always "1") Applied (Always "1") Applied (Always "1") Applied (Always "1") Applied (Always "1")

Output Range.Unit Index

Applied

Applied

Applied

Applied

Applied

Volumetric Flow Unit

Applied

Applied

Applied

Applied

Applied

Temperature Unit

Applied

Applied

Applied

Applied

Base Temperature

Applied

Applied

Pressure Unit

Applied

Applied

Base Pressure abs

Applied

Density Unit

Applied

Applied

Applied

Applied *1

Applied

Base Density

Applied *1

Applied

Deviation

Applied

First Temperature Coef.

Applied

Second Temperature Coef.

Applied

Applied: Target parameter is applied by DYFMVTool

*1:

Not applied when output unit is Normal or Standard