Autozero function – RADWAG WLC 0,6/A1/C/2 Precision Balance User Manual

Page 25

Advertising
background image

- 25 -

13.3. Autozero function

The autozero function has been implemented in order to assure precise

indications. This function controls and corrects

„0” indication. While the

function is active it compares the results continuously with constant

frequency. If two sequentional results differ less than the declared

value of autozero range, so the scale will be automatically zeroed

and the pictograms

and

will be displayed.

When AUTOZERO is disabled zero is not corrected automatically.

However, in particular cases, this function can disrupt the measurement

process e.g. slow pouring of liquid or powder on the weighing pan.

In this case, it is advisable to disable the autozero function.

Procedure:

Enter the submenu

<P1.rEAd> and then:

Fnnd no - filter disabled

Fnnd YES - filter enabled

Return to weighing:

See - 11.2.2.