Standard mass determined by its weighing – RADWAG WLC 0,6/A1/C/2 Precision Balance User Manual

Page 43

Advertising
background image

- 43 -

Return to weighing:

Press the

key twice.

16.5. Control of % deviation referring to the inscribed standard mass

Scale software enables control of deviation (in %) of weighed loads

mass referring to the inscribed standard mass. Mass of standard can

be determined by its weighing (

PrcA function) or entered to the scale

memory by an user (

PrcB function).

16.5.1. Standard mass determined by its weighing

Procedure:

Enter to

<PrcA> function:

You will see

<LoAd> on the display and then:

place an load on the pan which mass will be accepted as standard

press

to confirm this operating mode

after few seconds the indication

100,00% will be displayed

From this moment display will not indicate mass of weighed load

but deviation of load mass placed on the pan referring to the mass

of standard (in %).