Continuous transmission – RADWAG WLC 0,6/A1/C/2 Precision Balance User Manual

Page 72

Advertising
background image

- 72 -

sign

[space] for positive values or

[-] for negative values

mass

9 characters justified to the right

unit

3 characters justified to the left

command

3 characters justified to the left

Example 1:
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF –

the printout generated from

the scale after pressing ENTER/PRINT.

Example 2:
? _ - _ _ _ _ 2 . 2 3 7 _ l b _ CR LF -
the printout generated from

the scale after pressing ENTER/PRINT.

Example 3:
^ _ _ _ _ _ _ 0 . 0 0 0 _ k g _ CR LF
- the printout generated from

the scale after pressing ENTER/PRINT.

20.6. Continuous transmission

The indicator can work in a continuous transmission mode. It can

be switched on or off in parameters or using RS232 commands.

The frame format sent by the indicator in case of setting

<P2.Prnt> to

CntA:

1

2

3

4

5

6

7-15

16

17 18 19

20

21

S

I

space

stability

space

sign

mass

space

Unit

CR LF

Stability character

[space] if stable

[?] if not stable

[^] if an indication over the range

[v] if fan indication below the range

sign

[space] for positive values or

[-] for negative values

mass

9 characters justified to the right

unit

3 characters justified to the left

command

3 characters justified to the left