17.1.1. Manual internal calibration

Procedure:

While in weighing mode press

,

The scale program starts to check stability conditions for the

calibration process and displays the following message:

Then the program automatically goes to the internal calibration

procedure which is signalled by the following message:

After completion of the calibration process program returns

to the weighing mode,

Calibration process can be terminated anytime by pressing

which is signalled by the following message on the display:

Notice:

1. It should be remembered that internal calibration should

be performed with unloaded pan with keeping possibly

constant ambient conditions.

2. If the calibration process lasts longer than 15 seconds scale

software will react with

<Err8> displayed and a short sound

and then the calibration procedure will start again.