Automatic tare – RADWAG WLC 0,6/A1/C/2 Precision Balance User Manual
Page 45

- 45 -
Return to weighing:
Press the
key twice.
16.6. Automatic tare
This function is useful for fast net mass determination of weighed load
in case when tare value of is different for each load. In case when the
function is active the cycle of scales operating looks as follows:
press zeroing key when the pan is empty,
place the container for pieces,
when indication is stable
automatic tarring of the container mass will
be performed (
Net marker will appear in the upper part of the display),
place a sample into the package,
display will indicate net mass of sample,
remove the sample together with the container,
display will indicate tare mass with minus sign,
place a container for the next sample. When indication is stable
automatic tarring
will take place (Net marker will appear in the
upper part of the display),
place next sample into the package.
Procedure:
Return to weighing:
Press the
key twice.
- WLC 6/A2/C/2 Precision Balance WLC 1/A2/C/2 Precision Balance WLC 120/C2/K Precision Balance WLC 120/C2/R Precision Balance WLC 60/C2/K Precision Balance WLC 60/C2/R Precision Balance WLC 30/F1/R Precision Balance WLC 20/A2 Precision Balance WLC 12/F1/R Precision Balance WLC 10/A2 Precision Balance WLC 6/F1/K Precision Balance WLC 6/F1/R Precision Balance WLC 6/A2 Precision Balance WLC 2/A2 Precision Balance WLC 1/A2 Precision Balance WLC 0,6/A1 Precision Balance WLC 0,6/B1 Precision Balance