17. SCALE CALIBRATION

In precise scales changes of gravitational acceleration have noticeable

influence. The gravitational acceleration changes with altitude and

latitude. Every scale has to be adjusted to the place of use especially

when the place changes. Frequent calibration also prevents weighing

process from the influence of humidity and temperature.

For assuring the maximal accuracy of weighing a periodical user

calibration is required.

Calibration should be performed:

Before weighing process,

After a long break between series of measurements,

After the ambient temperature change.

Conditions of trigerring off calibration:

Automatic internal calibration:

Started by adequate temperature change,

Started after adjusted time period,

Started after powering up the device,

Manual internal calibration started from the keyboard,

Calibration with an external weight.

Caution:

Internal calibration is accessible only in WLC…/C/2 scales of WLC series.

In WLC…/C/2 scales calibration with an external weight is not accessible.

It should be remembered that the calibration process should be performed

with the empty pan! The calibration process can be terminated by

pressing

Esc when necessary.

17.1. Internal calibration

An option for WLC…/C/2 scales of WLC series only

The internal calibration process can be initiated manually or automatically.

Press

Cal to initiate it manually. Automatic calibration system performs

internal calibration and informs a user on the display about the course

of the process.