- 69 -

SU_A CR LF

SU_E CR LF

- command accepted and in progress

- timeout while waiting for stable results

SU_I CR LF

- command comprehended but cannot be executed

SU_A CR LF

MASS FRAME

- command accepted and in progress

- mass value in current unit is returned

Frame format:

1

2

3

4

5

6

7-15

16

17

18

19

20

21

S

U

space

stability

space

sign

mass

space

unit

CR

LF

Example:
S U CR LF
– computer command

S U _ A CR LF - command accepted and in progress

S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command done, mass

value in current unit is returned.

20.4.7. Send the result immediately in current unit

Syntax:

SUI CR LF

Possible answers:

SUI_I CR LF

- command comprehended but cannot be executed

MASS FRAME - mass value in current unit is returned immediately

Frame format:

1

2

3

4

5

6

7-15

16

17

18

19

20

21

S

U

I

stability

space

sign

mass

space

unit

CR

LF

Example:
S U I CR LF
– computer command

S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command executed

and mass returned