Send the result immediately in current unit, Frame format – RADWAG WLC 0,6/A1/C/2 Precision Balance User Manual
Page 69
- 69 -
SU_A CR LF
SU_E CR LF
- command accepted and in progress
- timeout while waiting for stable results
SU_I CR LF
- command comprehended but cannot be executed
SU_A CR LF
MASS FRAME
- command accepted and in progress
- mass value in current unit is returned
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
space
stability
space
sign
mass
space
unit
CR
LF
Example:
S U CR LF – computer command
S U _ A CR LF - command accepted and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command done, mass
value in current unit is returned.
20.4.7. Send the result immediately in current unit
Syntax:
SUI CR LF
Possible answers:
SUI_I CR LF
- command comprehended but cannot be executed
MASS FRAME - mass value in current unit is returned immediately
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
I
stability
space
sign
mass
space
unit
CR
LF
Example:
S U I CR LF – computer command
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command executed
and mass returned