Trouble shooting, Technical parameters, Precisions scales of wlc series – RADWAG WLC 0,6/A1/C/2 Precision Balance User Manual
Page 74
- 74 -
2. Error
FULL2 that appears on the display is also signalled by
a continuous sound until the cause of error disappears.
22. TROUBLE SHOOTING
Problem
Cause
Solution
Turning on does not
work
Discharged batteries.
Connect to mains or
change batteries
No batteries (not installed or
improperly installed)
Check the correctness of
installation (polarization)
The scale turns off
automatically
„t1” set to „YES” (Power
save)
In „othr” submenu change
„5.4 t1” to „no”
After turning on „LH”
message on the display
Loaded weight pan during
powering up
Unload the pan. Then the
scale will indicator zero.
23. TECHNICAL PARAMETERS
23.1. Precisions scales of WLC series
Scale type:
WLC 0,6/B1
WLC 1/A2
WLC 2/A2
-
-
Max capacity
0,6kg
1kg
2kg
Min capacity
0,5g
-
-
Reading division [d]
0,01g
0,01g
0,01g
Verification interval [e]
0,1g
-
-
Range of tare
-0,6kg
-1kg
-2kg
Repeatability
0,01g
0,03g
0,03g
Linearity
±0,01g
±0,03g
±0,03g
Pan size
125x145mm
195×195mm
Stabilization time
3 sec
Operation temperature
+15°C to +30°C
Ingress protection rating
IP43
Power supply
110÷230V AC 50÷60Hz / 11V AC and battery
Display
LCD (with backlight)
Supplied from batteries
35h (average time)
45h (average time)
Net / Gross weight
1,1/2kg
2,8/3,8kg
Package dimensions
320x210x150mm
440x280x190mm
- WLC 6/A2/C/2 Precision Balance WLC 1/A2/C/2 Precision Balance WLC 120/C2/K Precision Balance WLC 120/C2/R Precision Balance WLC 60/C2/K Precision Balance WLC 60/C2/R Precision Balance WLC 30/F1/R Precision Balance WLC 20/A2 Precision Balance WLC 12/F1/R Precision Balance WLC 10/A2 Precision Balance WLC 6/F1/K Precision Balance WLC 6/F1/R Precision Balance WLC 6/A2 Precision Balance WLC 2/A2 Precision Balance WLC 1/A2 Precision Balance WLC 0,6/A1 Precision Balance WLC 0,6/B1 Precision Balance