Send the result immediately in basic unit, Send the stable result in current unit, Frame format – RADWAG WLC 0,6/A1/C/2 Precision Balance User Manual

Page 68: Syntax: si cr lf, Possible answers

Advertising
background image

- 68 -

Frame format:

1

2-3

4

5

6

7-15

16

17

18

19

20

21

S

space

stability

space

sign

mass

space

unit

CR

LF

Example:
S CR LF
– computer command

S _ A CR LF - command accepted and in progress

S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF – command done,

mass value in basic unit is returned.

20.4.5. Send the result immediately in basic unit

Syntax:

SI CR LF

Possible answers:

SI_I CR LF

- command comprehended but cannot be executed at the

moment

MASS FRAME - mass value in basic unit is returned

Frame format:

1

2

3

4

5

6

7-15

16

17

18

19

20

21

S

I

space

stability

space

sign

mass

space

unit

CR

LF

Example:
S I CR LF
– computer command

S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command done, mass

value in basic unit is returned immediately.



20.4.6. Send the stable result in current unit

Syntax:

SU CR LF

Possible answers:

Advertising