- 56 -

17.1.2. Automatic internal calibration

The automatic calibration process can be triggered off by 3 different

factors:

Calibration after powering up

After performing the start procedure the scale program starts

to check stability conditions for the calibration process and

displays the following message:

Then the program automatically goes to the internal calibration

procedure which is signalled by the following message:

After completion of the calibration process program returns

to weighing mode.

Calibration triggered off by temperature changes

The scale has been equipped in the temperature monitoring

system;

Temperature triggers off calibration every time when the internal

system measures the temperature change greater than 3°C;

The calibration procedure triggered off by the temperature

change starts with checking which is signalled by the

following message: