Respond message format, Command’s description, Zeroing – RADWAG WLC 0,6/A1/C/2 Precision Balance User Manual
Page 66: Tarring

- 66 -
20.3. Respond message format
After sending a request message you can receive:
XX_A CR LF
command accepted and in progress
XX_D CR LF
command completed (appears only after XX_A)
XX_I CR LF
command comprehended but cannot be executed
XX _ ^ CR LF
command comprehended but time overflow error appeared
XX _ v CR LF
command comprehended but the indication below the
XX _ OK CR LF Command done
ES_CR LF
Command not comprehended
XX _ E CR LF
error while executing command – time limit for stable result
exceeded (limit time is a descriptive parameter of the scale)
XX - command name
_ - substitutes spaces
20.4. Command’s description
20.4.1. Zeroing
Syntax
Z CR LF
Possible answers:
Z_A CR LF
Z_D CR LF
- command accepted and in progress
- command completed
Z_A CR LF
Z_^ CR LF
- command accepted and in progress
- command comprehended but zero range overflow appeared
Z_A CR LF
Z_E CR LF
- command accepted and in progress
- time limit for stable result exceeded
Z_I CR LF
- command comprehended but cannot be executed
20.4.2. Tarring
Syntax:
T CR LF
Possible answers:
- WLC 6/A2/C/2 Precision Balance WLC 1/A2/C/2 Precision Balance WLC 120/C2/K Precision Balance WLC 120/C2/R Precision Balance WLC 60/C2/K Precision Balance WLC 60/C2/R Precision Balance WLC 30/F1/R Precision Balance WLC 20/A2 Precision Balance WLC 12/F1/R Precision Balance WLC 10/A2 Precision Balance WLC 6/F1/K Precision Balance WLC 6/F1/R Precision Balance WLC 6/A2 Precision Balance WLC 2/A2 Precision Balance WLC 1/A2 Precision Balance WLC 0,6/A1 Precision Balance WLC 0,6/B1 Precision Balance