Switch on continuous transmission in basic unit, Switch off continuous transmission in basic unit, Switch on continuous transmission in current unit – RADWAG WLC 0,6/A1/C/2 Precision Balance User Manual

Page 70: Syntax: c1 cr lf, Possible answers, Frame format, Syntax: c0 cr lf, Syntax: cu1 cr lf

Advertising
background image

- 70 -

20.4.8. Switch on continuous transmission in basic unit

Syntax:

C1 CR LF

Possible answers:

C1_I CR LF

- command comprehended but cannot be executed

C1_A CR LF

MASS FRAME

- command comprehended and in progress

- mass value in basic unit is returned

Frame format:

1

2

3

4

5

6

7-15

16

17

18

19

20

21

S

I

space

stability

space

sign

mass

space

unit

CR

LF

20.4.9. Switch off continuous transmission in basic unit

Syntax:

C0 CR LF

Possible answers:

C0_I CR LF

- command comprehended but cannot be executed

C0_A CR LF

- command comprehended and executed

20.4.10. Switch on continuous transmission in current unit

Syntax:

CU1 CR LF

Possible answers:

CU1_I CR LF

- command comprehended but cannot be executed

CU1_A CR LF

MASS FRAME

- command comprehended and in progress

- mass value in current unit is returned

Frame format:

1

2

3

4

5

6

7-15

16

17

18

19

20

21

S

U

I

stability

space

sign

mass

space

unit

CR

LF