INFICON IQM-233 Thin Film Deposition Controller PCI-Express Card Operating Manual User Manual

Page 53

Advertising
background image

3 - 13

PN

07

4-

58

4-

P1

A

IQM-233 Operating Manual

Sensor Fields

The Sensor 1, Sensor 2, and Sensor 3 fields correspond to the sensor(s)
connected to the SMA connectors on the IQM-233 card, with Sensor 1
corresponding to the SMA connector closest to the D-sub connector (refer to

Figure 2-7 on page 2-9

). Each sensor field contains the following parameters:

Density (g/cm

3

) . . . . . . . . . . . . . . . . 0.4 to 99.99 g/cm

3

Enter the density of the material to be deposited using the corresponding
sensor. See

Table A-1

for material density values. The default value is 1.00. If

the density value is not available for a material, see

section 6.2, Determining

Density, on page 6-1

.

Tooling (%) . . . . . . . . . . . . . . . . . . . 0.00 to 999.00

Enter a Tooling value to compensate for differences in the measured deposition
rate and actual deposition rate at the substrate due to the sensor or substrate
geometry (see

Figure 3-13

). The default value is 100.00. To determine the

Tooling value, see

section 6.3, Determining Tooling, on page 6-2

.

Figure 3-13 Tooling settings

Z-Ratio . . . . . . . . . . . . . . . . . . . . . . . 0.100 to 9.999

Enter a Z-Ratio value specific to the material being deposited to compensate
for the mechanical elasticity of the material to the quartz crystal. See

Table A-1

for material Z-Ratio values.

If the Z-Ratio of a material is not known, enter the default value of 1.000 for
Z-Ratio. Z-Ratio has a smaller effect when the crystal is relatively new. To
determine Z-Ratio experimentally, see

section 6.4, Determining Z-Ratio, on

page 6-3

.

NOTE: Click OK to save the Sensor Setup parameters.

Substrate

Substrate

Tooling

Over 100%

Tooling

Under 100%

Tooling over 100%

Tooling under 100%

Substrate

Substrate

Advertising