2.1.1.5 Ground Requirements

Low impedance cables or straps must be used to connect the chassis of all control
components to a common ground point on the vacuum chamber. The common
ground point must be connected to earth ground. See

section 2.1.1.5.1 on page 2-5

for the earth ground requirement.
Solid copper straps at least 12.7 mm (0.5 in.) wide and approximately
0.56 mm (0.022 in.) thick (as short as possible) are recommended where RF is
present. This is particularly important in high-noise e-beam systems.
See

Figure 2-2

for the recommended grounding method.

The oscillator is grounded to the IQM-233 card and crystal sensor through the
BNC cables and SMA/BNC adapter.

The crystal sensor is typically grounded to the wall of the vacuum system. If the
sensor feedthrough is not properly grounded to earth through the vacuum
system, connect a copper strap between the feedthrough and the common
ground point on the vacuum system.

WARNING - Risk Of Electric Shock

The computer in which the IQM-233 card is installed must
be connected to earth ground through a sealed
three-conductor power cable plugged into a socket outlet
with a protective ground terminal. The computer must
also be connected to the common ground point, shown
in

Figure 2-2

.

Extension cables must have three conductors including
a protective earth ground.

Figure 2-2 System grounding diagram

Earth

Vacuum System

Ground

Oscillator

Sensor

Feedthrough

Computer

Source

IQM-233

3-Conductor

Card

Copper Straps to

Common Ground

and Cables

Source Output

Cable

Power Cable

Power Supply