3.4 Technical Data

General

Approvals:

The instrument-specific marks are located next to
the identification label

Nominal voltage:

100 V - 120 V +/- 10%
230 V - 240 V +/- 10%

Nominal frequency:

50 to 60 Hz

Maximum power draw:

1400 VA

IEC 1010 classifications:

Protective class 1
Pollution degree 2
Overvoltage installation category II:

800 V impulse (120 V systems)

1500 V impulse (240 V systems)

Remote alarm relay:

24 V AC/DC, maximum 2 A
Connection:
potential-free contact
(can be operated both as normally open and as

normally closed circuit)

Primary fuses:

Fa. Schurter: type FST, mains power inlet 2 x T 8 A

Secondary fuses:

Melting fuse, Ø 6.3x32 mm
Type: Schurter FST
F1: T 2.0 A
F2: T 2.0 A
F3: T 2.5 A
F4: T 4.0 A

Operating temperature range:

15 °C to 40 °C

Relative humidity:

10% to 80%, non-condensing