3 tap water supply connection, Tap water supply connection, Setting up the instrument – Leica Biosystems ST5020 User Manual

Page 18

Advertising
background image

18

Instructions for Use, V1.9, RevD - 12/2012

4.3 Tap water supply connection

Fig. 6

3

Connecting the tap water inlet hose

Unpack the tap water inlet hose - mind the two gaskets (4).

4

5

6

Insert one gasket (4) into threaded fitting (6). Store away the remaining
gasket(s).

Insert strainer (3) into union nut (5) (see

Fig. 6

) which is screwed onto

the water tap. Do not insert any further gaskets, as the strainer acts
also as a seal.

7

7

4.

Setting up the instrument

Water filter

In addition to the filter sieve supplied (item 3 in

Fig. 6)

Leica recommends

installing a sediment filter between the water extraction outlet and
ST5020, to guarantee that the instrument is only operated with clean wa-
ter.

This sediment filter should have the following properties:

Filter fineness:

25 μm

Rate of flow:

approx. 10 l/min (with 5 wash stations)

Pressure constancy:

10 bar

Connection to the filter output: 1/2“ external thread

This filter insert must be checked and cleaned regularly.
If the filter insert is very dirty it must be replaced.