1 station buttons, 2 station types, Station buttons – Leica Biosystems ST5020 User Manual

Page 54: Station types, Operation

Advertising
background image

54

Instructions for Use, V1.9, RevD - 12/2012

5.5.2 Station types

Load station

•

There is a maximum of four load stations (L33-L36), L36 being non-
changeable. Load stations do not have a bar on the left border.

•

After a new rack has been placed into a load station, the second line of
the corresponding button indicates which program (color and code)
has been recognized and will be carried out.
The rhombus (#) indicates that the rack has been detected by the soft-
ware and will be processed as soon as possible.

•

The first line contains the station number (2) plus a letter (3)
indicating the station type.

These are:

R: Reagent station

L: Load station

O: Oven station

E: Unload station

W: Water station

Empty stations do not carry any letters.

•

At the left border of the button a bar (4) may be displayed,
providing different types of information, depending on the
type of station.

•

The second and third line are activated when a station is
used in a program run.

•

The first line of a station turns white (5) when the transfer
arm is located above that station.

•

In the second line the program code (6) is displayed in
the clip color that has been allocated.

•

The third line contains a rhombus (#) followed by the se-
quential number (7) of the rack. This rhombus counts the
number of racks which pass the station everyday. When
no. 99 is reached and everyday at 24:00, the rack counter
is reset to “0“.

5.5.1 Station buttons

Each of the 40 stations is configured as a button that can be pressed to directly access the
STATION DETAILS menu. The station buttons are divided into three lines containing differ-
ent types of information, depending on each individual configuration.

2

3

4

5

7

6

Fig. 51

5.

Operation

Advertising