1 station buttons, 2 station types, Station buttons – Leica Biosystems ST5020 User Manual

Page 54: Station types, Operation

Advertising
background image

54

Instructions for Use, V1.9, RevD - 12/2012

5.5.2 Station types

Load station

There is a maximum of four load stations (L33-L36), L36 being non-
changeable. Load stations do not have a bar on the left border.

After a new rack has been placed into a load station, the second line of
the corresponding button indicates which program (color and code)
has been recognized and will be carried out.
The rhombus (#) indicates that the rack has been detected by the soft-
ware and will be processed as soon as possible.

The first line contains the station number (2) plus a letter (3)
indicating the station type.

These are:

R: Reagent station

L: Load station

O: Oven station

E: Unload station

W: Water station

Empty stations do not carry any letters.

At the left border of the button a bar (4) may be displayed,
providing different types of information, depending on the
type of station.

The second and third line are activated when a station is
used in a program run.

The first line of a station turns white (5) when the transfer
arm is located above that station.

In the second line the program code (6) is displayed in
the clip color that has been allocated.

The third line contains a rhombus (#) followed by the se-
quential number (7) of the rack. This rhombus counts the
number of racks which pass the station everyday. When
no. 99 is reached and everyday at 24:00, the rack counter
is reset to “0“.

5.5.1 Station buttons

Each of the 40 stations is configured as a button that can be pressed to directly access the
STATION DETAILS menu. The station buttons are divided into three lines containing differ-
ent types of information, depending on each individual configuration.

2

3

4

5

7

6