Instructions for Use, V1.9, RevD - 12/2012

When pressing YES a second prompt opens up, where you need to
confirm that you have removed the rack from the station it was located
in last.
This step is necessary as the instrument itself is not able to detect
whether there is a rack in that station or not.

Wait until the transfer arm has finished the movement in
progress and is located in the parking position (rear left). Open
the lid and remove the rack. WAIT until the station number is
displayed on the screen - once this message is being dis-
played, confirm that the rack has been removed.

Press CONFIRM - the current staining run is aborted and removed from
the list of pending runs. Staining will be resumed starting with the next
rack.

If the program highlighted by the blue bar has already been
completed (rack already located in unload station or just being
transferred to unload station), the CANCEL RACK button is dis-
abled.

Cancelling racks

5.

Operation

Racks can be cancelled if needed.

Press STAINING SCHEDULE to access the stain-
ing schedule menu.

Select the rack you wish to cancel (highlight
with blue bar).

Press CANCEL RACK.

Press YES to confirm.
If NO is pressed, the instrument will switch back
to the staining schedule menu without any
changes being made.