5.4.9 Calibration dialog

Fig. 48

Functional test:

Calibrate the screen and verify the result ap-
proaching one of the screen buttons from all
four sides - if the screen has been successfully
calibrated, the button may not appear as pres-
sed (wider border) before you actually touch it.

In the CALIBRATION DIALOG menu you can en-
hance or restore touch screen response to but-
tons being pressed. We recommend to use a
pen for the calibration procedure.

3

2

1

Standard appearance

Button pressed

The pen must be placed exactly on the red dot
(1). The dot will move as shown in

Fig. 29

. To exit

the menu, press the dot once more in Position 4.

PRINT button:

Press for a printout of the
current station configu-
ration.

Instead of the graphical screen dis-
play, a list (

Fig. 47

) will be printed,

which, in addition to the information
provided by the graphical screen, also
shows the reagents allocated to each
station.

Fig. 47

5.

Operation